ACCURACY OF PRESSURE AND FLOW CAPACITIES OF FOUR ARTHROSCOPIC FLUID MANAGEMENT SYSTEMS

The present study was conducted to evaluate four different fluid delivery systems. A surgeon can trust all of the pumps when the pressure is set below 60 mm Hg.

Thomas Muellner, M.D.1, 2,Wolfgang A. Menth-Chiari, M.D.1, Lars Engebretsen, M.D., Ph.D.2, Roland Reihsner, Ph.D. 3, Josef Eberhardsteiner, Ph.D.3,

1 University of Vienna Medical School, Trauma Center, Austria 2 Oslo Orthopaedic University Clinic, Norway 3 Institute for Strength of Materials, Vienna University of Technology, Austria

Supported by a grant of the Austrian Society for Surgery and by the Austrian Society for Sports Medicine. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.

Running title: Arthroscopic fluid management systems

Key words: Arthroscopic fluid delivery systems-Infusion pumps-Pressure Characteristics-Flow Characteristics

INTRODUCTION

Good visualization is of key importance in arthroscopic surgery. In recent years, vast improvements have been made in the optical as well as the pump systems. Previous studies have reported improving mechanisms for pressure and flow control. In 1977, Gillquist et al. first reported a threshold pressure of 28 mm Hg as necessary for good visualization during knee arthroscopy.1 Ewing et al found that a pressure of 70 mm Hg was necessary for consistent and sustained capsular distension.2 Arangio and Kostelnik investigated the minimum adequate pressure required for arthroscopic diagnostic procedures and surgery of the knee.3 Their results showed that with an average pressure of 55 mm Hg, arthroscopy could be performed in all knee positions studied. Furthermore, a positive correlation between this pressure and the diastolic blood pressure was found. 3 During arthroscopy in a closed system, pressures in excess of 200 mm Hg can be generated in flexion and extension maneuvers.4, 5 Excessive fluid pressures may result in subcutaneous extravasal diffusion, leading to possible major complications.4-14

The aim of this study was to test four different fluid delivery systems used in routine arthroscopic surgery and to investigate the accuracy of their flow and pressure capacities.

MATERIALS AND METHODS

A knee model utilizing a thin-walled, hollow, spherical rubber cavity with a volume of 700 ml was used for this investigation. Two small holes were made into the ball and two commercially available cannulas (ø 8.25 mm, AR 6535, Arthrex®) were twisted into the ball to guarantee a watertight placement. The first was used for the outflow and to introduce a flat sensor (developed at the Vienna University of Technology) which consists of a pressure-measuring element, embedded in a capsule, working on the basis of the semiconductor technique. The hydrostatic surrounding of the measuring element allows a transformation of the inhomogeneous pressure engagements into a uniformly distributed pressure acting on the measuring element. Furthermore, the capsule protects the measuring element against intruding liquid. The pressure limit for the sensor is 5250 mmHg (7bar). The resolution of the sensor was 0.75 mmHg. The second cannula was used to introduce a 5mm wide-angle arthroscope and to direct the inflow through the metal sheath. 

Device representatives were asked to be responsible for the set-up of their pumps to guarantee the setup suggested by the manufacturer. The knee model and the devices tested were placed at the same height.

The investigation consisted of two parts: In the first part, the achieved pressure was measured with the pressure sensor within the ball and compared with the preset pressure values. From the lowest value the pressure was increased constantly while the flow was kept constant. In the second experimental part, the maximum flow per minute was tested using a beaker and a stopwatch. A measuring amplifier (DMC plus, HMB) and an iMac DV Computer (Apple Computer, Inc., Cupertino, CA, USA) were used for data storage and analysis.

Four different fluid management systems were investigated:

A)   Arthrex® AR-6450, (Arthrex Med. Instr. GmbH, Karlsfeld, Germany)

B)   Stryker® 1.5L High Flow Pump (Stryker Osteo GmbH, Muelheim, Germany)

C)   Arthro FMS® 4 (Future Medical Systems S.A. Geneva, Switzerland)

D)   InteliJet (Acufex ®, Smith+Nephew, Andover, MA, USA)

The Arthrex AR-6450 pump (A) allows flow rates of up to 1600 ml/min at pressure levels to 255 mm Hg. It has a one roller-pump and the pressure is controlled continuously with a sensor. The flow rate was set to 70 % and the pressure was increased from 40 mm Hg to 240 mm Hg in increments of 20 mm Hg according to the display on the device. Outflow was placed to gravity.

The Stryker® 1.5L High Flow Pump (B) allows flow rates of 100 to 1500 ml/min at pressures up to 200 mm Hg. It has a one-roller pump and three programs to save three different settings. With a flow rate of 1500 ml/min the pressure was increased at the same rate (20 mm Hg) from 40 to 160 mm Hg according to the device’s display. Higher pressures on the display could not be reached. Outflow was placed to gravity.

The Arthro FMS® 4 (C) pump is a computer-assisted dual pump system with two roller-pumps, providing precise control over both irrigation and suction. The integrated suction pump manages both shaver and cannula suction. The pump has a performance of 100 to 800 ml/min at pressure levels between 10 and 150. These pressure values represent fictitious numbers and are unitless. The investigation started at a pressure level of 10 and was increased by tens till the pressure level was 100 at a flow rate of 94 ml/min. According to the manufacturer, a value of 10 should be comparable to a pressure of 37 mm Hg, and a value of 150 corresponds to a pressure of 368 mm Hg.

The InteliJetpump (D) has a centrifugal pump, which allows flow rates in excess of 2000 ml/min. The outflow is controlled, conserving saline. Gravity inflow serves as a backup to the system. Pressure management is achieved through direct measurement at the cassette. The system is adjustable from 10 to 150 mm Hg. Using a flow of 30% to 40 %, the investigation started at a pressure of 40 mm Hg and was increased by increments of 20 mm Hg to a maximum of 150 mm Hg according to the device’s display. The flow was tested with the outflow placed to gravity, suction, and then the suggested original inflow cannula was used instead of the sheath for the arthroscope with the outflow placed to suction.

RESULTS

The highest measured pressures for the Arthrex® and the Stryker® pumps were 115 mm Hg (46 %) and 113 mm Hg (57 %), respectively. In the beginning, the measured values were almost in accordance with the values shown on the display of the devices. Pressures up to 60 mm Hg were measured adequately by the pressure sensors of the two devices. The higher the pressure was set on the device the greater the deviation found.

The highest measured pressures for the Arthro FMS® 4 and the InteliJet pumps were 251 mm Hg (101 %) and 132 mm Hg (88 %), respectively. Both devices measured the given pressure very accurately. While the Arthro FMS® 4 pump measured accurately throughout all ranges, the InteliJet pump prevailed only at lower pressure values. (Fig. 2)

The maximal flow values for the Arthrex® and Stryker® pumps were 900 ml/min (56.3 %) and 675 ml/min (45 %), respectively. The Arthro FMS® 4 and InteliJet pumps reached a maximal flow of 450 ml/min (56 %) and 1250 ml/min (62.5 %). With suction, the knee model collapsed despite full flow and maximum pressure using the InteliJet pump when the original inflow cannula was not used. By using the original inflow cannula the maximal flow per minute was increased only by 100 ml.

DISCUSSION

In 1989, Dolk and Augustini 15 compared three irrigation systems for motorized arthroscopic surgery. Using a joint model, pressure measurements were performed simultaneously within the chamber with a standard dome transducer and in the bag with a transducer-tipped catheter. With the Arthropump, intraarticular pressure was 30 to 35 mm Hg lower than the preset values at standard speed and a flow rate of 70 to 90 ml/min. With a higher flow rate and suction or shaver suction turned on, the pressure fell by 70 to 80 mm Hg compared with preset values. Also, the pressure recorded with the 3M arthroscopy pump was 10 to 20 mm Hg lower than the preset values. The aim of this study was to investigate whether present fluid management systems for arthroscopic surgery really reach the high-pressure values as suggested on their displays in an artificial setting. In our study, the experimental setup chosen was similar to that of routine arthroscopic procedures with regard to the flow. The manufacturers’ representatives suggested the respective flow rate usually necessary for good visualization during arthroscopy. While all other variables were kept constant, the pressure was increased continuously.

Considering the possible complications related to high pressures, such as synovial pouch rupture, fluid extravasation, and compartment syndrome, it is difficult to understand why all devices allow pressure values far above 100 mm Hg. Moreover, Sperber and Wredmark 16 investigated the capsular reaction to high intraarticular pressures. They concluded that in order to avoid capsular damage, knee arthroscopy should be done at intraarticular pressure levels below 120 mm Hg. 16 According to Burgaard et al, the intraarticular pressure should be as low as possible and should never exceed 150 mm Hg. 5 In this cadaver study 10 knees were filled at 35° of flexion to a pressure of 100 mm Hg, were subsequently extended and flexed, and all knees ruptured during this maneuver. 5 Also using a cadaveric model, Noyes and Spievack 17 measured the intraarticular and thigh pressures during arthroscopy. Intraarticular pressures ranging from 180 to 240 mm Hg produced synovial pouch rupture in all knees studied. 17 Conversely, in a clinical study, Ewing et al 2 measured a mean maximal pressure value of 268.5 mm Hg (±138.9; range 90-750 mm Hg) without observing massive extravasation or rupture of the synovial pouch. They explained that the differences between their results and those of Noyes and Spievack´s 17 could be explained by differences between clinical subjects and cadaveric specimens or by different pressurization rates. 2 

The two more sophisticated pump systems, the Arthro FMS® 4 and the InteliJet pumps, were able to reach the preset pressure. For the two other pumps, Arthrex® and Stryker®, large and small divergences between the preset and measured pressure were observed. These more simply constructed pumps never reached a pressure exceeding 135 mm Hg with the outflow cannula open. Both the Arthrex® and the Stryker® pumps had problems with the testing method chosen. In this study the reachable pressures when the outflow cannula was closed were not investigated, because in operative arthroscopic surgery a continuous flow is necessary to remove blood and debris. Considering the fact that an intraarticular pressure of below 120 mm Hg was suggested by Sperber and Wredmark 16 to avoid capsular damage, all pumps were able to reach this intraarticular pressure limit in our experimental setting.

With a positive pressure in the knee joint no compartmentation occurs within the suprapatellar pouch or anteromedial or posteromedial compartments. 18 Therefore, the hollow, spherical rubber cavity is a valid knee model, with one limitation left: no flexion or extension maneuvers, which are done frequently and at high-angle velocity during arthroscopy, can be performed. However, there is evidence in the literature, that with a closed system and even low baseline articular pressures about 20 mm Hg, intraarticular pressures in excess of 200 mm Hg are generated by flexing the knee. 4, 5 During routine arthroscopy rapid movements caused intraarticular pressure values up to 400 mm Hg. 15 

To our knowledge there are no reports about experimental flow measurements in the literature. Interestingly, none of the investigated pumps reached their published maximum flow values. The devices reached flow rates of 45 % to 62.5 % with respect to the published values of up to 2000 ml/min.

Flow through a vessel or tube follows Poiseuille´s law. The most important flow-regulating factor is the diameter of the inflow line. In our setting the flow was led through the arthroscope despite the high resistance in the spigot and water canal. But even with the use of the original inflow cannula the maximal flow of pump D could be increased only by 100 ml/min, from 1150 to 1250 ml/min.  

In conclusion, our investigation revealed that all four pumps are able to create pressure values considered by some authors to be high enough to damage the capsule of the knee. The two simpler pumps do not create pressures accurately and as high as set on the display, perhaps because these pumps are not able to control the outflow. In contrast, the two outflow-controlling pumps create pressures in a precise manner. None of them reached the maximum published flow values in our experimental setup. In the simpler pumps the pressure on the display must be set to higher values compared with the outflow-controlling pumps, to achieve the same intraarticular pressure. Below a displayed pressure of 60 mm Hg a surgeon can trust the arthroscopic pumps used. When higher pressures are set, a surgeon who is familiar only with simple roller pumps must be aware that outflow-controlled pumps, such as the ones tested in this study, reach the preset pressure very accurately with the possibility of capsular damage. The authors do not understand why manufacturers allow their pumps to generate pressure values - at least on the display - beyond the limits suggested in the literature and that fictitious, unitless numbers are used by one of them.  

ACKNOWLEDGEMENT The authors thank Josef Beer, Ph.D., University of Technology Vienna, Institute for Strength of Materials, Vienna, Austria for his technical assistance and experienced help, and Faith McLellan, Ph.D., Philadelphia PA, USA, for editorial review of this material.

REFERENCES

1.         Gillquist, J., Hagberg, G., Oretorp, N. Arthroscopy in acute injuries of the knee joint. Acta Orthop Scand 1977; 482: 190-196

2.         Ewing, J. W., Noe, D. A., Kitaoka, H. B., Askew, M. J. Intra-articular pressures during arthroscopic knee surgery. Arthroscopy 1986; 24: 264-269

3.         Arangio, G., Kostelnik, K. E. Intraarticular pressures in a gravity-fed arthroscopy fluid delivery system. Arthroscopy 1992; 83: 341-344

4.         Funk, D. A., Noyes, F. R., Grood, E. S., Hoffman, S. D. Effect of flexion angle on the pressure-volume of the human knee. Arthroscopy 1991; 71: 86-90

5.         Burgaard, P., Blyme, P. J., Olsen, P. M., Kristensen, G.  Rupture of the knee capsule from articular hyperpressure. Experiments in cadaver knees. Acta Orthop Scand 1988; 596: 692-69

6.         Abbushi, W., Egbert, R., Pichelmeier, R., Kovacs, J. Compartment syndrome of the forearm due to infusion and transfusion using a pressure pump. Anasthesiol Intensivmed Notfallmed Schmerzther 1991; 266: 348-351

7.         Bomberg, B. C., Hurley, P. E., Clark, C. A., McLaughlin, C. S. Complications associated with the use of an infusion pump during knee arthroscopy. Arthroscopy 1992; 82: 224-228

8.         Carpenter, J. F. Complications associated with the use of an infusion pump during knee arthroscopy. Arthroscopy 1993; 92: 244

9.         Ekman, E. F., Poehling, G. G. An experimental assessment of the risk of compartment syndrome during knee arthroscopy. Arthroscopy 1996; 122: 193-199

10.       Bartlett, C. S., DiFelice, G. S., Buly, R. L., Quinn, T. J., Green, D. S., Helfet, D. L. Cardiac arrest as a result of intraabdominal extravasation of fluid during arthroscopic removal of a loose body from the hip joint of a patient with an acetabular fracture. J Orthop Trauma 1998; 124: 294-299

11.       Belanger, M., Fadale, P. Compartment syndrome of the leg after arthroscopic examination of a tibial plateau fracture. Case report and review of the literature. Arthroscopy 1997; 135: 646-651

12.       Seiler, J. G. I., Valadie, A. L. I., Drvaric, D. M., Frederick , R. W., Whitesides, T. E. J. Perioperative compartment syndrome. A report of four cases. J Bone Joint Surg 1996;  78 (Am)4: 600-602

13.       Sorrentino, F., Eggli, S., Stricker, U., Ballmer, F. T., Hertel, R. Missed compartment syndrome after anterior cruciate ligament-plasty following continous peridural anesthesia. Unfallchirurg 1998; 1016: 491-494

14.       Romero, J., Smit, C. M., Zanetti, M. Massive intraperitoneal and extraperitoneal accumulation of irrigation fluid as a complication during knee arthroscopy. Arthroscopy 1998; 144: 401-404

15.       Dolk, T., Augustini, B. G. Three irrigation systems for motorized arthroscopic surgery: a comparative experimental and clinical study. Arthroscopy 1989; 54: 307-314

16.       Sperber, A., Wredmark, T. Tensile properties of the knee-joint capsule at an elevated intraarticular pressure. Acta Orthop Scand 1998; 695: 484-488

17.       Noyes, F. R., Spievack, E. S. Extraarticular fluid dissection in tissues during arthroscopy. A report of clinical cases and a study of intraarticular and thigh pressures in cadavers. Am J Sports Med 1982; 106: 346-351

18.       Sperber, A., Wredmark, T. Multicompartmental pressures in the knee joint during arthroscopy. Arthroscopy 1993; 95: 566-569

 

TABLES AND FIGURES

Table 1

 

Arthrex®

Stryker®

Arthro FMS® 4

InteliJet

Pressure (mm Hg)

 

 

 

 

Published

255 mm Hg

200 mm Hg

100

150 mm Hg

Shown on Display

240 mm Hg

160 mm Hg

100

150 mm Hg

Measured

115 mm Hg

113 mm Hg

251 mm Hg

132 mm Hg

%

46

57

101

88

Max. Flow (ml/min)

 

 

 

 

Published

1600

1500

800

2000

Measured

900

675

450

1250

%

56

45

56

63

 

Both the Stryker® and the Arthrex® pumps were far away from the published maximum pressures, while the two other pumps were almost accurate.

Comparison of the four devices with regard to the reached pressure.

 

 

 

Dokumente

Muellner_Pumpe.pdf
Format: PDF
Größe: 537,35 kB

Publikationen

Beliefs and Attitudes of Members of the ESSKA Regarding the Treatment of Hamstring Injuries

IMG_1053.jpg Significant variation is seen in clinical opinion and decision-making regarding hamstring injuries among members of the ESSKA. Areas of significant clinical uncertainty should be the focus of future research and medical education for orthopaedic surgeons who treat hamstring injuries.
» Mehr

 

„Schlüssel zum aktiven Alter“

IMG_1051.JPG Univ. Doz. Dr. Thomas Müllner, NÖTV-Verbandsarzt und Facharzt für Unfallchirurgie und Orthopädie, über den Funktionsverlust „alter“ Muskulatur, Sport als Alternative zur Hightech-Medizin und weshalb ein 70-Jähriger die gleiche Kraft haben kann wie 26-Jähriger.
» Mehr

 

Schmerzen in der Hüfte

2011.jpg AMIS® Der anteriore Zugang folgt den Prinzipien von MIS:kleinere Hautschnitte keine beeinträchtigen Muskeln und/oder Sehnen (Basis 1/2011)
» Mehr

 

Das Kreuz mit dem Band

4.jpg Das Knie wird unter anderem vom vorderen Kreuzband stabilisiert. Reißt das Kreuzband, ist die Stabilität oft dahin - ein Schreckensszenario für viele sportlich aktive Menschen. NetDoktor.at befragte dazu Univ. Doz. Dr. Thomas Müllner, PhD.
» Mehr

 

Herbert Köberl

Ohne Titel.jpg Herbert Köberl hat den Eingriff gut überstanden. Nach der AMIS Hüft Prothese konnte bereits 18 Stunden nach der Operation ohne Krücken gehen. Seit dem 5. postoerativen Tag fährt er Auto.
» Mehr